OMD-15
General Details
OMD-15 is designed for the special requirements on boiler water and condensate field of monitor type was: An examination by Germanischer Lloyd, taking into account the “Technical regulations for steam boilers TRD 604”, Part 1 and in accordance with the provisions of IMO Resolution MEPC.60 (33) (Test Specifications for Ölgehaltsmeßgeräte in the shipping sector) have been successfully completed and the relevant certificates are us before. The benefits of the OMD-15 Series are, by using the latest technology and selected components, essentially:
- reliability
- easy handling
- improved solid suppression
- adjustable parameters
- compact device
- LED display
- and light weight (3.5kg)
Additional accessories such as sample preparation units and pumps available on request.
Variants:
OMD-15 without automatic cell cleaning
OMD-15A with automatic cell cleaning
Technical Details
Eigenschaft | Wert |
---|---|
Messbereich | 0 – 9,9 ppm ungelöstes Öl |
Genauigkeit | Besser als +/- 1 ppm (bei bekannter Ölsorte) |
Gehäuse Schutzart | IP 65 |
Abmessungen OMD-15 | 175 x 190 x 220 mm (B x H x T)) |
Abmessungen OMD-15 A | 210 x 330 x 220 mm (B x H x T) |
Probenwasser Temperatur | +1° bis +90° C |
Alarmkreise 1+2 | Einstellbar zwischen 1 – 9,9 ppm |
Zeitverzögerung der Alarmkreise | Einstellbar zwischen 2- 30 Sekunden |
Netzanschluß | 24 Volt AC/DC, andere Spannungen lieferbar |
Ausgangssignal | 0 – 20 mA oder 4 – 20 mA, einstellbar |